Simulador de paciente de electrocardiografía: diseño y desarrollo
| dc.contributor.advisor | Sierra Bueno, Daniel Alfonso | |
| dc.contributor.advisor | Quintero Muñoz, Jorge Eduardo | |
| dc.contributor.author | Álvarez Jimenez, Roberth Arturo | |
| dc.contributor.author | Bravo Pérez, Juan Camilo | |
| dc.date.accessioned | 2024-03-03T22:06:43Z | |
| dc.date.available | 2015 | |
| dc.date.available | 2024-03-03T22:06:43Z | |
| dc.date.created | 2015 | |
| dc.date.issued | 2015 | |
| dc.description.abstract | Este proyecto presenta el diseño e implementación de un simulador de paciente de electrocardiografía que permite simular las doce derivaciones cardiacas de un paciente en condiciones patológicas y en condiciones normales, esto con _x001C_nes netamente académicos. El equipo cuenta con una LCD donde se muestra información sobre la señal que se está simulando, un teclado para seleccionar la señal deseada y un microcontrolador. Para el microcontrolador se usó una tarjeta de desarrollo de Texas Instruments que cumple con las demandas del proyecto que son memoria _x001D_ash y número de GPIOs (General Purpose Input/Output). Se utilizó la tarjeta TivaTM C Series TM4C123G LaunchPad. Luego de esto, se procedió a diseñar el método para capturar las señales que iban a ir almacenadas en la memoria del microcontrolador. Para esto se utilizó un simulador Flyke Biomedical MedSim300B Patient Simulator, del cual se capturaron las señales mediante un ampli_x001C_cador de instrumentación de ganancia 100, un _x001C_ltro pasabajas de frecuencia de corte de 40Hz y un _x001C_ltro pasa altas de frecuencia de corte de 0,5Hz conectados en cascada. Luego del _x001C_ltrado, se utilizó una DAQ de National Instruments para capturar la señal de forma discretizada. Mediante el software de Labwiew se _x001C_ltró la señal nuevamente con _x001C_ltros digitales de las mismas características de los _x001C_ltros analógicos y adicionalmente se le agregó un _x001C_ltro rechaza-banda. Finalmente se programó el microcontrolador y se acoplaron los diferentes periféricos para dejar funcionando el equipo. Se espera que este equipo sea utilizado como ayuda para proyectos que requieran señales cardiacas de una persona para _x001C_nes académicos. | |
| dc.description.abstractenglish | This work presents the design and implementation of a patient's ECG simulator that pretends to simulate the twelve ECG-lead of a patient in pathological conditions and under normal conditions, only for academic purposes. Initially, it was decided that the device would have an LCD with information about the signal that was pretending to be shown, a keyboard to select the desired signal and a microcontroller. For the microcontroller, a launchpad from Texas Instruments was selected, since it meets the demands of the project that basically were _x001D_ash memory and number of GPIOs (General Purpose Input / Output); in particular, it was decided to use the Tiva TM4C123G LaunchPad TM C Series. After that, we proceeded to design a method to capture the signals that were going to been stored in the microcontroller memory. For this, was used a professional Fluke patient simulator, in which signals were captured by a 100 gain instrumentation ampli_x001C_er, a low pass _x001C_lter cuto_x001B_ frequency of 40Hz and a high pass _x001C_lter cuto_x001B_ frequency of 0.5 Hz cascade. After _x001C_ltering, a DAQ from National Instruments was used to capture the signal in discretized form. Using LabView software the signals were _x001C_ltered again with digital _x001C_lters with the same characteristics than the analog _x001C_lters already used, and further added a bandstop _x001C_lter. Finally the algorithm was implemented in the microcontroller and the externals peripherals were coupled together. It is expected that this device is used as an aid for projects that need an ECG patient signal for academic purpose. | |
| dc.description.degreelevel | Pregrado | |
| dc.description.degreename | Ingeniero Electrónico | |
| dc.format.mimetype | application/pdf | |
| dc.identifier.instname | Universidad Industrial de Santander | |
| dc.identifier.reponame | Universidad Industrial de Santander | |
| dc.identifier.repourl | https://noesis.uis.edu.co | |
| dc.identifier.uri | https://noesis.uis.edu.co/handle/20.500.14071/32551 | |
| dc.language.iso | spa | |
| dc.publisher | Universidad Industrial de Santander | |
| dc.publisher.faculty | Facultad de Ingenierías Fisicomecánicas | |
| dc.publisher.program | Ingeniería Electrónica | |
| dc.publisher.school | Escuela de Ingenierías Eléctrica, Electrónica y Telecomunicaciones | |
| dc.rights | http://creativecommons.org/licenses/by/4.0/ | |
| dc.rights.accessrights | info:eu-repo/semantics/openAccess | |
| dc.rights.creativecommons | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
| dc.rights.license | Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) | |
| dc.rights.uri | http://creativecommons.org/licenses/by-nc/4.0 | |
| dc.subject | Derivaciones Cardiacas | |
| dc.subject | Condiciones Patológicas | |
| dc.subject | Microcontrolador | |
| dc.subject | Software | |
| dc.subject | Lcd | |
| dc.subject | Filtros Digitales | |
| dc.subject | Filtros Análogos | |
| dc.subject.keyword | Ecg-Lead | |
| dc.subject.keyword | Pathological Conditions | |
| dc.subject.keyword | Microcontroller | |
| dc.subject.keyword | Software | |
| dc.subject.keyword | Lcd | |
| dc.subject.keyword | Digital _x001C_Lters | |
| dc.subject.keyword | Analog _x001C_Lters | |
| dc.title | Simulador de paciente de electrocardiografía: diseño y desarrollo | |
| dc.title.english | Simulador de paciente de electrocardiografía - diseño y desarrollo | |
| dc.type.coar | http://purl.org/coar/version/c_b1a7d7d4d402bcce | |
| dc.type.hasversion | http://purl.org/coar/resource_type/c_7a1f | |
| dc.type.local | Tesis/Trabajo de grado - Monografía - Pregrado |