Publicación: Implementación de un método primario para la calibración de un patrón tipo pistón para medición de bajos caudales de gas, propiedad del CDT de GAS
| dc.contributor.advisor | Fuentes Osorio, José Augusto | |
| dc.contributor.author | Calle Tafur, Eices Milena | |
| dc.contributor.evaluator | García Navas, Edinson Odilio | |
| dc.contributor.evaluator | Prieto Jiménez, Natalia | |
| dc.date.accessioned | 2025-10-28T19:57:45Z | |
| dc.date.available | 2025-10-28T19:57:45Z | |
| dc.date.created | 2025-08-25 | |
| dc.date.embargoEnd | 2030-08-25 | |
| dc.date.issued | 2025-08-25 | |
| dc.description.abstract | El presente trabajo de aplicación propone la implementación del método gravimétrico como alternativa para la calibración del patrón primario tipo pistón, empleado en la evaluación de desempeño de medidores en bajos caudales de gas (0,001 dm³/min a 20 dm³/min). Su aplicación es particularmente relevante, en sectores relacionados con la industria de hidrocarburos gaseosos como es el caso de análisis de calidad de gases, el control de fugas, dilución de gases y la confiabilidad en medición de flujo de pruebas piloto. Actualmente, el CDT de GAS realiza la calibración de este patrón mediante un método dimensional el cual cuenta con acreditación en la ISO/IEC 17025. El presente estudio plantea un método complementario basado en la medición de masa, con el propósito de mejorar la trazabilidad metrológica y la confiabilidad en la medición del flujo de gas. El estudio se desarrolló en tres fases. Primero, se diseñó e integró el proceso de calibración complementario basado en el método gravimétrico, seleccionando los equipos y la metodología que permitiera lograr el nivel de exactitud similar o mejor al método dimensional actualmente utilizado (Incertidumbre expandida: ±8,8 Pulsos/dm3). Luego, se llevó a cabo la caracterización metrológica del método, cuantificando la incertidumbre mediante el método GUM y simulación Monte Carlo. Finalmente, se realizó un análisis estadístico de error normalizado (En) sobre los resultados experimentales para evidenciar la comparabilidad entre ambos métodos. Como resultado final, se evidenció que el valor de En fue de 0,06 (el límite permisible es En ≤1). Este resultado permite confirmar que no existen diferencias significativas en términos de precisión y exactitud entre el método de calibración dimensional y el gravimétrico. Finalmente, se confirmó que la constante de calibración del pistón (Kp) es 11895,3 ± 4,3 pulsos/dm3. | |
| dc.description.abstractenglish | This application proposes the implementation of the gravimetric method as an alternative for calibrating the primary piston-type standard, used to evaluate the performance of meters at low gas flow rates (0.001 dm³/min to 20 dm³/min). Its application is particularly relevant in sectors related to the gaseous hydrocarbon industry, such as gas quality analysis, leak control, gas dilution, and reliability in pilot test flow measurements. Currently, the GAS CDT calibrates this standard using a dimensional method accredited by ISO/IEC 17025. This study proposes a complementary method based on mass measurement, with the aim of improving metrological traceability and reliability in gas flow measurement. The study was carried out in three phases. First, the complementary calibration process based on the gravimetric method was designed and integrated, selecting the equipment and methodology that would achieve a level of accuracy similar to or better than the currently used dimensional method (expanded uncertainty: ±8.8 pulses/dm3). Then, the metrological characterization of the method was carried out, quantifying the uncertainty using the GUM method and Monte Carlo simulation. Finally, a statistical analysis of normalized error (En) was performed on the experimental results to demonstrate comparability between both methods. The final result was that the En value was 0.06 (the permissible limit is En ≤1). This result confirms that there are no significant differences in terms of precision and accuracy between the dimensional and gravimetric calibration methods. Finally, the piston calibration constant (Kp) was confirmed to be 11895.3 ± 4.3 pulses/dm3. | |
| dc.description.degreelevel | Maestría | |
| dc.description.degreename | Magíster en Ingeniería de Petróleos y Gas | |
| dc.format.mimetype | application/pdf | |
| dc.identifier.instname | Universidad Industrial de Santander | |
| dc.identifier.reponame | Universidad Industrial de Santander | |
| dc.identifier.repourl | https://noesis.uis.edu.co | |
| dc.identifier.uri | https://noesis.uis.edu.co/handle/20.500.14071/46266 | |
| dc.language.iso | spa | |
| dc.publisher | Universidad Industrial de Santander | |
| dc.publisher.faculty | Facultad de Ingeníerias Fisicoquímicas | |
| dc.publisher.program | Maestría en Ingeniería de Petróleos y Gas | |
| dc.publisher.school | Escuela de Ingeniería de Petróleos | |
| dc.rights | info:eu-repo/semantics/embargoedAccess | |
| dc.rights.accessrights | info:eu-repo/semantics/openAccess | |
| dc.rights.coar | http://purl.org/coar/access_right/c_f1cf | |
| dc.rights.creativecommons | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
| dc.rights.license | Atribución-NoComercial-SinDerivadas 2.5 Colombia (CC BY-NC-ND 2.5 CO) | |
| dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | |
| dc.subject | Medición de flujo de gas | |
| dc.subject | Patrón primario | |
| dc.subject | Calibración gravimétrica | |
| dc.subject | Trazabilidad metrológica | |
| dc.subject.keyword | Gas Flow Measurement | |
| dc.subject.keyword | Primary Standard | |
| dc.subject.keyword | Gravimetric Calibration | |
| dc.subject.keyword | Metrological Traceability | |
| dc.title | Implementación de un método primario para la calibración de un patrón tipo pistón para medición de bajos caudales de gas, propiedad del CDT de GAS | |
| dc.title.english | Implementation of a primary method for calibrating a piston-type standard for low-flow gas measurement, owned by the CDT de Gas | |
| dc.type.coar | http://purl.org/coar/resource_type/c_bdcc | |
| dc.type.hasversion | http://purl.org/coar/version/c_b1a7d7d4d402bcce | |
| dc.type.local | Tesis/Trabajo de grado - Monografía - Maestría | |
| dspace.entity.type | Publication |
Archivos
Bloque original
1 - 4 de 4
Cargando...
- Nombre:
- Carta autorización.pdf
- Tamaño:
- 360.23 KB
- Formato:
- Adobe Portable Document Format
Cargando...
- Nombre:
- Nota de proyecto.pdf
- Tamaño:
- 115.37 KB
- Formato:
- Adobe Portable Document Format
Cargando...
- Nombre:
- Carta de confidencialidad.pdf
- Tamaño:
- 278.39 KB
- Formato:
- Adobe Portable Document Format
Bloque de licencias
1 - 1 de 1
Cargando...
- Nombre:
- license.txt
- Tamaño:
- 2.18 KB
- Formato:
- Item-specific license agreed to upon submission
- Descripción: