Publicación: Disminución en la incertidumbre de un patrón tipo vasos comunicantes para calibración de presión, con un alcance manométrico de -5,2 kpa a 5,2 kpa
| dc.contributor.advisor | García Sánchez, Carlos Eduardo | |
| dc.contributor.advisor | Angulo, Jhon Alejandro | |
| dc.contributor.author | Prieto Rojas, Jairo Alberto | |
| dc.contributor.author | Rojas Gomez, Cristian Eduardo | |
| dc.date.accessioned | 2024-03-04T00:08:43Z | |
| dc.date.available | 2018 | |
| dc.date.available | 2024-03-04T00:08:43Z | |
| dc.date.created | 2018 | |
| dc.date.issued | 2018 | |
| dc.description.abstract | La demanda insatisfecha en Colombia, respecto a la calibración de bajas presiones, se debe principalmente a la insuficiencia de patrones secundarios, que brinden baja incertidumbre y cuenten con certificación ONAC. Esta investigación tuvo como propósito, tomar el patrón de vasos comunicantes Dwyer 1425 Series Hook Gage del laboratorio CDT de Gas y modificarlo, de tal manera que su incertidumbre estimada por el método GUM, sea menor a la original y por tanto se habilite este para realizar calibraciones. Para cumplir tal objetivo, primero se hizo una estimación de incertidumbre en el patrón original, para identificar las fuentes de variabilidad que más aportan duda en las mediciones. Se obtuvo que la medición de nivel en los vasos es la fuente más relevante, debido a su falta de precisión. En segundo lugar, se diseñaron 4 propuestas enfocadas a una disminución en el aporte por las mediciones de nivel, para posteriormente escoger una de estas propuestas en base al costo y a la incertidumbre. Se encontró que la mejor opción es suprimir la medición de nivel en un vaso, e incorporar mejoras ópticas en el otro, para continuar lecturas de nivel con micrómetro. Por último, se compararon las diferencias de incertidumbre entre el patrón original, el modificado y la incertidumbre requerida, de allí se evidencio una brecha significativa entre la incertidumbre del patrón original y modificado. Además, se logró un cumplimiento en los requerimientos del laboratorio en un rango manométrico de 2.5 kPa a 5.2 kPa, en presión y vacío. | |
| dc.description.abstractenglish | The unsatisfied demand in Colombia, regarding the calibration of low pressures, is mainly due to the lack of secondary patterns that provide low uncertainty and have ONAC certification. The purpose of this research is to take the Dwyer 1425 Series Hook Gage communicating vessel pattern from the Gas CDT laboratory and modify it, in such a way that its uncertainty estimated by the GUM method is lower than the original one and therefore it is enabled to perform calibrations. To accomplish this aim, the uncertainty in the original pattern is first estimated to identify the sources of variability that most contribute to doubt in the measurements, from which it was obtained that the level measurement in the vessels is the most relevant source, due to its lack of precision. Secondly, 4 proposals focused on a decrease in the contribution by the level measurements were designed, to later choose one of these proposals based on the cost and the uncertainty, from which it was obtained that the best option is to suppress the measurement level in a vessel and incorporate optical improvements in the other to continue level measurements with micrometer. Finally, the differences of uncertainty between the original, modified pattern and the required uncertainty were compared, from there a significant gap was obtained between the original and modified uncertainty of pattern, in addition a compliance in the laboratory requirements was achieved in a manometric range of 2.5 kPa to 5.2 kPa, in pressure and vacuum. | |
| dc.description.degreelevel | Pregrado | |
| dc.description.degreename | Ingeniero Químico | |
| dc.format.mimetype | application/pdf | |
| dc.identifier.instname | Universidad Industrial de Santander | |
| dc.identifier.reponame | Universidad Industrial de Santander | |
| dc.identifier.repourl | https://noesis.uis.edu.co | |
| dc.identifier.uri | https://noesis.uis.edu.co/handle/20.500.14071/38969 | |
| dc.language.iso | spa | |
| dc.publisher | Universidad Industrial de Santander | |
| dc.publisher.faculty | Facultad de Ingenierías Fisicoquímicas | |
| dc.publisher.program | Ingeniería Química | |
| dc.publisher.school | Escuela de Ingeniería Química | |
| dc.rights | http://creativecommons.org/licenses/by/4.0/ | |
| dc.rights.accessrights | info:eu-repo/semantics/openAccess | |
| dc.rights.creativecommons | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
| dc.rights.license | Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) | |
| dc.rights.uri | http://creativecommons.org/licenses/by-nc/4.0 | |
| dc.subject | Vasos Comunicantes | |
| dc.subject | Incertidumbre | |
| dc.subject | Fuente De Variabilidad | |
| dc.subject | Aporte De Incertidumbre | |
| dc.subject | Método Gum | |
| dc.subject | Precisión. | |
| dc.subject.keyword | Communicating Vessels | |
| dc.subject.keyword | Uncertainty | |
| dc.subject.keyword | Source Of Variability | |
| dc.subject.keyword | Contribution Of Uncertainty | |
| dc.subject.keyword | Gum Method | |
| dc.subject.keyword | Accuracy. | |
| dc.title | Disminución en la incertidumbre de un patrón tipo vasos comunicantes para calibración de presión, con un alcance manométrico de -5,2 kpa a 5,2 kpa | |
| dc.title.english | Decrease of the uncertainty of a communicating vessel pattern for pressure calibration, with a manometric range of -5.2 kpa to 5.2 kpa* | |
| dc.type.coar | http://purl.org/coar/version/c_b1a7d7d4d402bcce | |
| dc.type.hasversion | http://purl.org/coar/resource_type/c_7a1f | |
| dc.type.local | Tesis/Trabajo de grado - Monografía - Pregrado | |
| dspace.entity.type | Publication |
Archivos
Bloque original
1 - 3 de 3
Cargando...
- Nombre:
- Carta de autorización.pdf
- Tamaño:
- 358.47 KB
- Formato:
- Adobe Portable Document Format
Cargando...
- Nombre:
- Nota de proyecto.pdf
- Tamaño:
- 239.96 KB
- Formato:
- Adobe Portable Document Format