Práctica empresarial en ACIEND S.A.S como apoyo a la evaluación de problemas en la calibración de equipos de Ultrasonido (Phased Array) e implementación de técnica por C-scan para inspección de soldaduras y daños por corrosión

dc.contributor.advisorGalán Pinilla, Carlos Andrés
dc.contributor.authorMuñoz Ríos, Maure Loraine
dc.contributor.evaluatorGonzález Hernández, Andrés Giovanni
dc.contributor.evaluatorPérez Ceballos, Ana María
dc.date.accessioned2022-04-20T21:16:33Z
dc.date.available2022-04-20T21:16:33Z
dc.date.created2022-03-30
dc.date.embargoEnd2027-03-30
dc.date.issued2022-03-30
dc.description.abstractLa técnica de Ensayo No Destructivo (END) de Ultrasonido por Phased Array o arreglo de fases, es ampliamente utilizada en la industria petroquímica, para evaluar la calidad de juntas soldadas y accesorios de tuberías, entre otros componentes. Está técnica avanzada de inspección hace parte de los servicios prestados por la empresa ACIEND SAS ubicada en Barrancabermeja-Santander. Además cuenta con otras técnicas convencionales de inspección. Con el desarrollo de la presente práctica empresarial, se estudiaron fallas reportadas con respecto a los procesos de calibración de los equipos de Ultrasonido, con el fin de identificar las variables o parámetros que provocan la incorrecta calibración. Para esto se evaluaron a los tres inspectores que prestan sus servicios de forma permanente en ACIEND S.A.S, de lo cual, se evidenció que existen daños de Software, deterioro de las zapatas y por mejorar habilidad por parte de un inspector. Como resultado de la caracterización de equipos se identificó que la empresa puede ampliar su portafolio de servicios en la identificación de corrosión por medio de la aplicación de C-scan, para lo cual se requiere realizar una inversión en nuevos palpadores y zapatas, con la finalidad de efectuar la técnica a los diámetros de servicio más comunes ofertados en la empresa (3, 6 y 8 pulg). Finalmente, la preparación obtenida durante la práctica mediante capacitaciones, realización de certificaciones y ejecución de los ensayos en campo, permitieron la participación en diversos proyectos, entre ellos, el proyecto de evaluación de integridad del gasoducto Payoa-Galán de la empresa Petrosantander, en el que participé durante la etapa de inspección, análisis y discusión del caso de estudio, en el cual, se identificaron diversos mecanismos de daño mecánicos y por corrosión; los cuales fueron evaluados bajo norma y son reportados en el presente informe.
dc.description.abstractenglishThe Phased Array Ultrasound Non-Destructive Testing (NDT) technique is widely used in the petrochemical industry to assess the quality of welded joints and pipe fittings, among other components. This advanced inspection technique is part of the services provided by the company ACIEND SAS located in Barrancabermeja-Santander, in addition to other conventional inspection techniques. When developing the business practice, it was evidenced that the company had flaws with respect to the calibration processes of the Ultrasound equipment, for which the basic variables that caused the incorrect calibration were identified; evaluated to the three inspectors who provide their services in ACIEND S.A.S, of which, Software damage, deterioration of the shoes and skill on the part of an inspector were evidenced. As a result of the characterization of equipment, it was identified that the company can expand its portfolio of services in the identification of corrosion through the application of C-scan, for which it is required to make an investment in new probes and shoes, with the purpose to carry out the technique to the most common service diameters offered in the company (3, 6 and 8 inches). Finally, the preparation obtained during the internship through training, certifications and execution of field tests, allowed participation in various projects, including the Petrosantander company's Payoa-Galán gas pipeline integrity assessment project, in the that I participated during the inspection, analysis and discussion stage of the case study, in which various mechanical and corrosion damage mechanisms were identified; which were evaluated under the standard and are reported in this report.
dc.description.degreelevelPregrado
dc.description.degreenameIngeniero Metalúrgico
dc.format.mimetypeapplication/pdf
dc.identifier.instnameUniversidad Industrial de Santander
dc.identifier.reponameUniversidad Industrial de Santander
dc.identifier.repourlhttps://noesis.uis.edu.co
dc.identifier.urihttps://noesis.uis.edu.co/handle/20.500.14071/10089
dc.language.isospa
dc.publisherUniversidad Industrial de Santander
dc.publisher.facultyFacultad de Ingeníerias Fisicoquímicas
dc.publisher.programIngeniería Metalúrgica
dc.publisher.schoolEscuela de Ingeniería Metalúrgica y Ciencia de Materiales
dc.rightsinfo:eu-repo/semantics/openAccess
dc.rights.accessrightsinfo:eu-repo/semantics/openAccess
dc.rights.coarhttp://purl.org/coar/access_right/c_abf2
dc.rights.creativecommonsAtribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0)
dc.rights.licenseAttribution-NonCommercial 4.0 International (CC BY-NC 4.0)
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subjectAciend SAS
dc.subjectEnsayos no Destructivos
dc.subjectUltrasonido Phased Array
dc.subjectCalibración
dc.subjectC-scan
dc.subjectIndustria petroquímica
dc.subject.keywordACIEND S.A.S
dc.subject.keywordNon-Destructive Testing
dc.subject.keywordUltrasound Phased Arrays
dc.subject.keywordcalibration
dc.subject.keywordC-scan
dc.subject.keywordpetrochemical industry
dc.titlePráctica empresarial en ACIEND S.A.S como apoyo a la evaluación de problemas en la calibración de equipos de Ultrasonido (Phased Array) e implementación de técnica por C-scan para inspección de soldaduras y daños por corrosión
dc.title.englishBusiness internship at ACIEND S.A.S as support for the evaluation of problems in the calibration of Ultrasound equipment (Phased Array) and implementation of C-scan technique for inspection of welds and corrosion damage
dc.type.coarhttp://purl.org/coar/resource_type/c_7a1f
dc.type.hasversionhttp://purl.org/coar/version/c_b1a7d7d4d402bcce
dc.type.localTesis/Trabajo de grado - Monografía - Pregrado
dspace.entity.type
Files
Original bundle
Now showing 1 - 4 of 4
No Thumbnail Available
Name:
Documento.pdf
Size:
1.3 MB
Format:
Adobe Portable Document Format
No Thumbnail Available
Name:
Carta de autorización.pdf
Size:
205.92 KB
Format:
Adobe Portable Document Format
Description:
No Thumbnail Available
Name:
Carta de confidencialidad.pdf
Size:
136.1 KB
Format:
Adobe Portable Document Format
Description:
No Thumbnail Available
Name:
Nota de proyecto.pdf
Size:
179.84 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
2.18 KB
Format:
Item-specific license agreed to upon submission
Description: