Publicación: Diseño de un medidor de conductividad eléctrica de soluciones salinas, haciendo uso de una celda de conductancia calculable, para calibración de una sonda tetrapolar utilizada en la medición de impedancia eléctrica en tejido humano
| dc.contributor.advisor | Barrero Perez, Jaime Guillermo | |
| dc.contributor.author | Jaimes Morales, Samuel Alberto | |
| dc.contributor.author | Valencia Quintero, Juan Pablo | |
| dc.contributor.author | Miranda Mercado, David Alejandro | |
| dc.date.accessioned | 2024-03-03T16:06:39Z | |
| dc.date.available | 2006 | |
| dc.date.available | 2024-03-03T16:06:39Z | |
| dc.date.created | 2006 | |
| dc.date.issued | 2006 | |
| dc.description.abstract | En el presente trabajo se diseñó e implementó un medidor de conductividad eléctrica para solución salina, el cual usa el método de Van der Pauw y una celda calculable tipo Morón. El factor de celda para este tipo de celda es constante y sólo depende de su geometría, por lo tanto, puede ser utilizada como estándar primario. El rango de medición del instrumento es de 80 mS/cm a 5200 mS/cm, valores de conductividad eléctrica de interés en el estudio de detección temprana de cáncer. Para estudiar las propiedades eléctricas de los tejidos y detectar precozmente el cáncer, es necesario utilizar una sonda de medición, por ejemplo, una sonda tetrapolar. Muchas sondas de medición no son calculables, por lo tanto, necesitan ser calibradas con una celda de estándar primario. Los estudios que se están haciendo en la Universidad Industrial de Santander para la detección de cáncer de cuello uterino están basados en una sonda de medición tetrapolar que necesita ser calibrada. El método de Van der Pauw necesita la conmutación de los electrodos de corriente (dos electrodos) y electrodos de medición de voltaje (dos electrodos). Las señales de corriente fueron implementadas con un generador de pulsos, un filtro pasa bajas y una fuente de corriente Howland, y la medición de voltaje con un amplificador de instrumentación y un medidor de RMS real. Un multiplexor analógico fue utilizado para la conmutación, y el proceso de control del dispositivo fue hecho con un microcontrolador de ocho bits de Motorota. El dispositivo diseñado fue probado con soluciones salinas, NaCl, de diferentes concentraciones. Estas soluciones fueron obtenidas diluyendo NaCl al 0.9% en agua destilada. Se obtuvieron medidas de conductividad eléctrica con una incertidumbre de 1.602% y una desviación estándar de 1.644%, en el peor de los casos. | |
| dc.description.abstractenglish | It was development an electric conductivity meter for saline solutions, it use Van der Pauw's method and a calculable Moron type conductance cell. The form factor for this cell is constant and it is function of its geometry, then, it cell will be used as a primary standard. The measurement range of the instrument is 80 mS/cm to 5200 mS/cm, the resistivities values of interests in the study of early cancer screening. To study tissue properties, and early cancer screening, is necessary used a measurement probe, e.g. a tetrapolar probe. Many measurement probe are not calculable cell, for that, it is necessary calibrate it whit a primary standard cell. The studies, for early cancer screening, are doing in the Universidad Industrial de Santander are based in a tetrapolar measurement probe that need be calibrate. The van der Pauw method needs the commutation of the current signal electrodes (two electrodes) and voltage measurement electrodes (two electrodes). The current signal was implemented whit a pulse wave generator, a low pass filter and a Howland current source, and, the voltage measurement whit an instrumentation amplifier and a true RMS converter. An analog multiplexer was used to do the commutation, and the control process whit a Motorola eight bits microcontroller. The development device was tested with different concentrations saline solutions, NaCl. Its solutions was obtained to become diluted NaCl to 0.9% in distilled water. The accuracy obtained was ±1.602% whit a standard deviation ±1.644%, in the worse case. | |
| dc.description.degreelevel | Pregrado | |
| dc.description.degreename | Ingeniero Electrónico | |
| dc.format.mimetype | application/pdf | |
| dc.identifier.instname | Universidad Industrial de Santander | |
| dc.identifier.reponame | Universidad Industrial de Santander | |
| dc.identifier.repourl | https://noesis.uis.edu.co | |
| dc.identifier.uri | https://noesis.uis.edu.co/handle/20.500.14071/18398 | |
| dc.language.iso | spa | |
| dc.publisher | Universidad Industrial de Santander | |
| dc.publisher.faculty | Facultad de Ingenierías Fisicomecánicas | |
| dc.publisher.program | Ingeniería Electrónica | |
| dc.publisher.school | Escuela de Ingenierías Eléctrica, Electrónica y Telecomunicaciones | |
| dc.rights | http://creativecommons.org/licenses/by/4.0/ | |
| dc.rights.accessrights | info:eu-repo/semantics/openAccess | |
| dc.rights.creativecommons | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
| dc.rights.license | Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) | |
| dc.rights.uri | http://creativecommons.org/licenses/by-nc/4.0 | |
| dc.subject | Conductividad eléctrica | |
| dc.subject | Soluciones salinas | |
| dc.subject | Van der Pauw | |
| dc.subject | Celdas de conductancia | |
| dc.subject | Estándar primario | |
| dc.subject | estándar secundario | |
| dc.subject | factor de forma. | |
| dc.subject.keyword | Electric conductivity | |
| dc.subject.keyword | Saline solutions | |
| dc.subject.keyword | Van der Pauw | |
| dc.subject.keyword | Conductance cells | |
| dc.subject.keyword | Primary standard | |
| dc.subject.keyword | secondary standard | |
| dc.subject.keyword | form factor. | |
| dc.title | Diseño de un medidor de conductividad eléctrica de soluciones salinas, haciendo uso de una celda de conductancia calculable, para calibración de una sonda tetrapolar utilizada en la medición de impedancia eléctrica en tejido humano | |
| dc.title.english | Design of an electric conductivity meter for saline solutions, using a calculable conductance cell, to the calibration of a tetrapolar measurement probe used in measurements of electric bioimpedance | |
| dc.type.coar | http://purl.org/coar/version/c_b1a7d7d4d402bcce | |
| dc.type.hasversion | http://purl.org/coar/resource_type/c_7a1f | |
| dc.type.local | Tesis/Trabajo de grado - Monografía - Pregrado | |
| dspace.entity.type | Publication |