Modelo funcional de un equipo generador de ondas ultrasónicas para fisioterapia : diseño y construcción
dc.contributor.advisor | Villamizar Mejia, Rodolfo | |
dc.contributor.advisor | Bautista, Ramiro Alfonso | |
dc.contributor.author | Castillo Cárdenas, Sonia Milena | |
dc.contributor.author | Cristancho Cuervo, Jorge Humberto | |
dc.contributor.author | Cristiano Rodriguez, Julian efren | |
dc.date.accessioned | 2024-03-03T16:33:26Z | |
dc.date.available | 2007 | |
dc.date.available | 2024-03-03T16:33:26Z | |
dc.date.created | 2007 | |
dc.date.issued | 2007 | |
dc.description.abstract | El ultrasonido es usado como agente terapéutico para tratar diversas afecciones físicas, aprovechando los efectos que causa en el organismo; estas ondas son generadas por transductores piezoeléctricos a partir de una señal periódica proveniente de un generador de ondas ultrasónicas, denominándose al conjunto completo generador-transductor ‚equipo de terapia por ultrasonido™. En la actualidad estos equipos permiten la variación de cuatro parámetros importantes: tiempo de tratamiento, intensidad de potencia, frecuencia de onda y modalidad de aplicación de la onda (continua o pulsátil). Para el mercado colombiano el costo de los equipos usados para llevar a cabo el tratamiento descrito es elevado, lo cual restringe el fácil acceso al mismo. En este trabajo se propone el diseño y la construcción de un modelo funcional electrónico capaz de generar ondas ultrasónicas mediante la programación de los parámetros mencionados. El equipo final basado en el modelo funcional será utilizado en fisioterapia para el tratamiento de las patologías que requieren de calor profundo, con un costo accesible en el país. Los resultados del presente trabajo muestran que el modelo funcional diseñado y construido es capaz de generar una onda ultrasónica senoidal con baja distorsión armónica, con la potencia de salida, frecuencia, modalidad de aplicación, y tiempo de terapia programados, dentro de los márgenes de tolerancia estipulados en las normas pertinentes y con los diseños considerados; además se contribuye al desarrollo de equipos de ultrasonido en la región, a partir de la transferencia y adaptación de la tecnología existente para los mismos. El diseño se basó en las normas establecidas para equipos de terapia por ultrasonido por la Comisión Electrotécnica Internacional, IEC (International Electrothecnical Commission) y recopilada por el Instituto Colombiano de Normas Técnicas y Certificación, ICONTEC. | |
dc.description.abstractenglish | Ultrasound is used as a therapeutic agent to treat diverse physical affections, taking advantage of the effects that it causes in the organism; these waves are generated by piezoelectric transducers from an periodic signal originated by an ultrasonic waves generator, denominating to the complete generator-transducer set 'therapy equipment by ultrasound'. At present time these equipments allows the variation of four important parameters: treatment time, power intensity, wave frequency and wave application modality (continuous or pulsed). For Colombian market, the used equipment to carry out the described treatment is expensive, which restricts its easy access. This work exposes the design and construction of an electronic functional model able to generate ultrasonic waves by means of the programming of the mentioned parameters. The final equipment based on the functional model will be used in physical therapy to treat pathologies that require of deep heat, with an accessible cost in the country. Results of the present work show that the designed and constructed functional model is able to generate a sine ultrasonic wave with low harmonic distortion, with programmed power output, frequency, application modality, and time of therapy, within the stipulated tolerance limits in the pertinent standards and with the considered designs; in addition, this work contributes to the development of ultrasound equipments in the region, starting from the transference and adaptation of the existing technology for such. The design was based on the established standards for therapy equipments by ultrasound by the IEC (International Electrothecnical Commission) and compiled by the Colombian Institute of Technical Standards and Certification, ICONTEC (Instituto Colombiano de Normas Técnicas y Certificación). | |
dc.description.degreelevel | Pregrado | |
dc.description.degreename | Ingeniero Electrónico | |
dc.format.mimetype | application/pdf | |
dc.identifier.instname | Universidad Industrial de Santander | |
dc.identifier.reponame | Universidad Industrial de Santander | |
dc.identifier.repourl | https://noesis.uis.edu.co | |
dc.identifier.uri | https://noesis.uis.edu.co/handle/20.500.14071/19782 | |
dc.language.iso | spa | |
dc.publisher | Universidad Industrial de Santander | |
dc.publisher.faculty | Facultad de Ingenierías Fisicomecánicas | |
dc.publisher.program | Ingeniería Electrónica | |
dc.publisher.school | Escuela de Ingenierías Eléctrica, Electrónica y Telecomunicaciones | |
dc.rights | http://creativecommons.org/licenses/by/4.0/ | |
dc.rights.accessrights | info:eu-repo/semantics/openAccess | |
dc.rights.creativecommons | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
dc.rights.license | Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc/4.0 | |
dc.subject | Aplicaciones Ultrasónicas | |
dc.subject | Conversión De Potencia | |
dc.subject | Generador De Señales | |
dc.subject | Realimentación | |
dc.subject | Terapia | |
dc.subject | Transductor | |
dc.subject.keyword | Ultrasonic Applications | |
dc.subject.keyword | Power Conversion | |
dc.subject.keyword | Signal Generator | |
dc.subject.keyword | Feedback | |
dc.subject.keyword | Therapy | |
dc.subject.keyword | Piezoelectric | |
dc.title | Modelo funcional de un equipo generador de ondas ultrasónicas para fisioterapia : diseño y construcción | |
dc.title.english | Functional model of an ultrasonic waves generating equipment for physical therapy: design and | |
dc.type.coar | http://purl.org/coar/version/c_b1a7d7d4d402bcce | |
dc.type.hasversion | http://purl.org/coar/resource_type/c_7a1f | |
dc.type.local | Tesis/Trabajo de grado - Monografía - Pregrado |