Diseño de un medidor de espejo enfriado para la verificación de calidad en el gas natural con base en el límite sobre el punto de rocío establecido por el RUT
dc.contributor.advisor | García Sánchez, Gabriel Fernando | |
dc.contributor.advisor | García Sánchez, Carlos Eduardo | |
dc.contributor.author | Morantes Rodríguez, Édgar Mauricio | |
dc.contributor.evaluator | Quiroga Méndez, Jabid Eduardo | |
dc.contributor.evaluator | Chacón Velasco, Jorge Luis | |
dc.date.accessioned | 2022-11-23T12:36:36Z | |
dc.date.available | 2022-11-23T12:36:36Z | |
dc.date.created | 2022-11-13 | |
dc.date.issued | 2022-11-13 | |
dc.description.abstract | La Comisión de Regulación de Energía y Gas (CREG) mediante el Reglamento Único de Transporte de gas natural (RUT) estipuló especificaciones de calidad del gas natural entregado al transportador por el agente en la entrada del sistema de transporte y por el transportador en el punto de salida, con el propósito de facilitar el intercambio de gases sin cambiar la seguridad operativa, eficiencia y desempeño. Por parte de las entidades encargadas de transportar el gas deben asegurar que el punto de rocío de hidrocarburos para cualquier presión no supere el valor de 45°F (7.2°C). Para ello se utiliza un instrumento llamado medidor de espejo enfriado, el cual se identifica como un equipo sujeto a errores de inspección visual por parte del operario, ya que el encargado genera una disminución manual de temperatura en el espejo hasta identificar de manera objetiva el inicio de condensación en el gas natural. Por tanto, se diseñó un prototipo de medidor de espejo enfriado con detección automática basado en el principio Dark Spot, el cual cuenta con un par de sensores barrera encargados de identificar la condensación en el gas natural, además de un controlador PID ajustado para mantener la temperatura del espejo constante a 7.2°C con el fin de cumplir el requisito establecido en el RUT. Para verificar su funcionamiento se realizó una simulación en OpenFoam con los parámetros reales seleccionados, donde permite concluir que existe una mejora en el manejo e interpretación de datos frente a los equipos manuales existentes actualmente. | |
dc.description.abstractenglish | The Energy and Gas Regulatory Commission (CREG) through the Single Regulation for the Transportation of Natural Gas (RUT) stipulated quality specifications for natural gas delivered to the transporter by the agent at the entrance of the transportation system and by the transporter at the exit point, with the purpose of facilitating the exchange of gases without changing the operational safety, efficiency and performance. On the part of the entities in charge of transporting the gas, they must ensure that the hydrocarbon dew point for any pressure does not exceed a value of 45°F (7.2°C). For this purpose, an instrument called a chilled mirror meter is used, which is identified as an equipment subject to visual inspection errors by the operator, since the person in charge generates a manual decrease of temperature in the mirror until objectively identifying the beginning of condensation in the natural gas. Therefore, a prototype of a cooled mirror meter with automatic detection based on the Dark Spot principle was designed, which has a pair of barrier sensors in charge of identifying condensation in the natural gas, in addition to a PID controller adjusted to maintain the mirror temperature constant at 7.2°C in order to meet the requirement established in the RUT. To verify its operation, a simulation was carried out in OpenFoam with the real parameters selected, where it was concluded that there is an improvement in data handling and interpretation compared to the currently existing manual equipment. | |
dc.description.degreelevel | Pregrado | |
dc.description.degreename | Ingeniero Mecánico | |
dc.format.mimetype | application/pdf | |
dc.identifier.instname | Universidad Industrial de Santander | |
dc.identifier.reponame | Universidad Industrial de Santander | |
dc.identifier.repourl | https://noesis.uis.edu.co | |
dc.identifier.uri | https://noesis.uis.edu.co/handle/20.500.14071/12207 | |
dc.language.iso | spa | |
dc.publisher | Universidad Industrial de Santander | |
dc.publisher.faculty | Facultad de Ingeníerias Fisicomecánicas | |
dc.publisher.program | Ingeniería Mecánica | |
dc.publisher.school | Escuela de Ingeniería Mecánica | |
dc.rights | info:eu-repo/semantics/openAccess | |
dc.rights.accessrights | info:eu-repo/semantics/openAccess | |
dc.rights.coar | http://purl.org/coar/access_right/c_abf2 | |
dc.rights.creativecommons | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
dc.rights.license | Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | |
dc.subject | Punto de rocío | |
dc.subject | Hidrocarburos | |
dc.subject | Detección automática | |
dc.subject | OpenFoam | |
dc.subject.keyword | Dew point | |
dc.subject.keyword | Hydrocarbons | |
dc.subject.keyword | Automatic Detection | |
dc.subject.keyword | OpenFoam | |
dc.title | Diseño de un medidor de espejo enfriado para la verificación de calidad en el gas natural con base en el límite sobre el punto de rocío establecido por el RUT | |
dc.title.english | Design of a chilled mirror meter for gas quality verification based on the limit on the dew point established by the RUT | |
dc.type.coar | http://purl.org/coar/resource_type/c_7a1f | |
dc.type.hasversion | http://purl.org/coar/version/c_b1a7d7d4d402bcce | |
dc.type.local | Tesis/Trabajo de grado - Monografía - Pregrado | |
dspace.entity.type |
Files
Original bundle
1 - 3 of 3
No Thumbnail Available
- Name:
- Carta de Autorización.pdf
- Size:
- 143.22 KB
- Format:
- Adobe Portable Document Format
No Thumbnail Available
- Name:
- Nota de Proyecto.pdf
- Size:
- 820.03 KB
- Format:
- Adobe Portable Document Format
License bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- license.txt
- Size:
- 2.18 KB
- Format:
- Item-specific license agreed to upon submission
- Description: