dc.description.abstract | En el presente trabajo se aborda un estudio fluidodinámico de dos configuraciones de una cánula de salida para un sistema de asistencia ventricular, compuesta por un sistema de impulsión de flujo axial, HeartMate II LVAD. Inicialmente se hace una investigación acerca del sistema circulatorio y de mayor interés la arteria aorta, por lo que, se acude a profesionales en el área médica, se consulta en la clínica Fundación Cardiovascular de Colombia fcv en Bucaramanga Santander, con el fin de encontrar información que permita tener claridad sobre la geometría de la cánula y la arteria aorta, se obtiene como respuesta del instituto un examen médico que contiene imágenes obtenidas por Tomografía Axial Computada (TAC) las cuales son leídas en la plataforma DICOM, estas imágenes son llevadas a un software computacional Mimics Medical 21.0, el cual con este se realiza un modelado tridimensional de la arteria aorta ascendente, luego de ello se realiza las construcción de la cánula en el software Solid Works y el ensamble cánula-aorta ascendente, al tener el sólido completo es importado a la plataforma de ANSYS FLUENT (CFD) para así realizar la simulación numérica a partir del método de volúmenes finitos. Con el estudio fluidodinámico se busca analizar el comportamiento de la sangre en la zona donde se intercepta la cánula de salida y la aorta ascendente, especialmente los flujos provenientes del sistema pulsátil y el ventrículo izquierdo, por lo que, en las dos configuraciones se desea evaluar qué diferencias presentan el flujo sanguíneo como son las zonas de baja velocidad, distribuciones de cortes y líneas de corriente. Como resultado de este estudio, se logró observar que el ángulo comprendido entre la cánula y la aorta ascendente tiene relación, ya que, este al ser menor evita fenómenos y patrones indeseados en el flujo, dando así una proyección a la hora de la intervención quirúrgica. * Trabajo de grado ** Facultad de Ingenierías Físico-Mecánicas. Escuela de Ingeniería Mecánica- Director: Manuel de Jesús Martínez, Doctorado en Ingeniería Mecánica. | |
dc.description.abstractenglish | The present project is about a fluid dynamic study of two configurations of an outlet cannula for a ventricular assist system, composed of an axial flow impulsion system, HeartMate II LVAD. Initially, the investigation is made about the circulatory system and the aorta artery, reason why, professionals in the medical area were consulted in the clinic Cardiovascular Foundation of Colombia (FCV) in Bucaramanga Santander, in order to find information that allows to have clarity about the geometry of the cannula and the aorta artery, it is obtained as a response of the institute, a medical examination that contains images obtained by Computed Axial Tomography (TAC) which are read in the DICOM platform, these images are taken to a computational software Mimics Medical 21.0, which with this a three-dimensional modeling of the ascending aorta artery is made, after that the construction of the cannula is made in the SolidWorks software and the ascending cannula-aorta assembly, after having this, the complete solid is imported to the ANSYS FLUENT (CFD) platform to make the numerical simulation from the Finite Volume Method. The fluid dynamic study seeks to analyze the behavior of the blood in the area where the exit cannula and the ascending aorta are intercepted, especially the flows from the pulsatile system and the left ventricle. Therefore, in the two configurations we want to evaluate the differences in blood flow such as low speed areas, wall shear stress and current lines. As a result of this study, it was possible to observe that the angle between the cannula and the ascending aorta is related, since this being smaller, it avoids unwanted phenomena and patterns in the flow, thus giving a projection at the time of surgery. * | |
dc.description.degreelevel | Pregrado | |
dc.description.degreename | Ingeniero Mecánico | |
dc.format.mimetype | application/pdf | |
dc.identifier.instname | Universidad Industrial de Santander | |
dc.identifier.reponame | Universidad Industrial de Santander | |
dc.identifier.repourl | https://noesis.uis.edu.co | |
dc.identifier.uri | https://noesis.uis.edu.co/handle/20.500.14071/13660 | |
dc.language.iso | spa | |
dc.publisher | Universidad Industrial de Santander | |
dc.publisher.faculty | Facultad de Ingenierías Fisicomecánicas | |
dc.publisher.program | Ingeniería Mecánica | |
dc.publisher.school | Escuela de Ingeniería Mecánica | |
dc.rights | info:eu-repo/semantics/openAccess | |
dc.rights.accessrights | info:eu-repo/semantics/openAccess | |
dc.rights.coar | http://purl.org/coar/access_right/c_abf2 | |
dc.rights.creativecommons | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
dc.rights.license | Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | |
dc.subject | Cánula | |
dc.subject | Aorta Ascendente | |
dc.subject | Bomba De Flujo Axial | |
dc.subject | Simulación Numérica De Fluidos | |
dc.subject | Método De Volúmenes Finitos. | |
dc.subject.keyword | Cannula | |
dc.subject.keyword | Ascending Aorta | |
dc.subject.keyword | Axial Flow Pump | |
dc.subject.keyword | Numerical Fluid Simulation | |
dc.subject.keyword | Finite Volume Method. | |
dc.title | Estudio fluidodinamico de una canula de salida del soporte mecanico circulatorio (smc) heartmate ® ii lvad mediante el metodo de volumenes finitos | |
dc.title.english | Fluidodynamic study of the outlet cannula o from the circulatory mechanical support (smc) using the finite volume method to understand the physical changes in the blood flow* | |
dc.type.coar | http://purl.org/coar/version/c_b1a7d7d4d402bcce | |
dc.type.hasversion | http://purl.org/coar/resource_type/c_7a1f | |
dc.type.local | Tesis/Trabajo de grado - Monografía - Pregrado | |
dspace.entity.type |
Files
Original bundle
1 - 3 of 3
No Thumbnail Available
- Name:
- Carta de autorización.pdf
- Size:
- 607.19 KB
- Format:
- Adobe Portable Document Format
No Thumbnail Available
- Name:
- Nota de proyecto.pdf
- Size:
- 243.92 KB
- Format:
- Adobe Portable Document Format