Evaluación del Efecto del Uso de Aire Atmosférico, Aire Sintético y la Actualización del Modelado Matemático de las Propiedades del Aire Involucradas Sobre la Calibración de los Principales Medidores de Volumen y Flujo de Gas de la Corporación Centro de Desarrollo Tecnológico del Gas.
dc.contributor.advisor | García Sánchez, Gabriel Fernando | |
dc.contributor.advisor | García Sánchez, Carlos Eduardo | |
dc.contributor.author | Sarmiento Palomino, Luis Miguel | |
dc.contributor.evaluator | Quiroga Méndez, Jabid Eduardo | |
dc.contributor.evaluator | Chacón Velasco, Jorge Luis | |
dc.date.accessioned | 2023-08-09T14:48:08Z | |
dc.date.available | 2023-08-09T14:48:08Z | |
dc.date.created | 2023-08-01 | |
dc.date.issued | 2023-08-01 | |
dc.description.abstract | En este trabajo se realiza una evaluación del efecto de la humedad y la variación en la proporción de nitrógeno y oxígeno del aire sintético, en los procesos de calibración de medidores de volumen y flujo de gas de la corporación CDT de Gas. La variación de estos parámetros de composición se debe al comportamiento del sistema de secado y el método de preparación del aire sintético. Se identifican la densidad y viscosidad como únicas propiedades que intervienen en el proceso de calibración, estando la viscosidad presente solo en la calibración con elementos de flujo laminar (LFE). Se realizan pruebas de secado de aire comprimido con equipo desecante regenerativo, y se diagnóstica y repara su falla obteniendo humedades de tan solo 15 ± 6 ppmv. Se compara el valor de la densidad y viscosidad a las mismas condiciones para distintas composiciones del aire. Son identificados los modelos actualmente empleados por el CDT para el cálculo de la densidad y viscosidad del aire seco y húmedo, además de proponer nuevos modelos para las propiedades de estas composiciones y para el aire sintético. Los datos de referencia para las propiedades de las distintas composiciones de aire son obtenidos mediante el software REFPROP en su versión 10,0. Cómo producto de este análisis, se encuentra que la composición con la máxima desviación respecto al aire atmosférico estándar es el aire sintético con máximo contenido de nitrógeno, resultando en un 0.6% mayor en el valor del caudal en LFE. También se realiza la comprobación experimental del efecto de la composición al emplear aire secado por el equipo desecante y aire sintético, en calibraciones entre patrón laminar y medidor térmico másico, así como medidor de pistón reciprocante. En estás calibraciones se emplea el puntaje En como criterio estadístico para establecer si los errores relativos del caudal en los distintos tipos de aire son comparables entre si. | |
dc.description.abstractenglish | This work contains an evaluation of the effect of humidity and the nitrogen-oxygen proportion variation for synthetic air over the calibration of gas volume and flow meters of the CDT de Gas corporation. The variation of these composition parameters is related to the behavior of the drying system and the method for preparing synthetic air. Density and viscosity are identified as the only properties involved in the calibration process, viscosity being present only in the calibration with laminar flow elements (LFE). Compressed air-drying tests are carried out in desiccant regenerative dryer, apart from diagnosing and repair a failure achieving a humidity of just 15 ± 6 ppmv. The values of density and viscosity are compared at the same conditions for different air compositions. In the same way, the relative error in the value of the flow rate obtained in LFE is calculated by introducing the value of the properties of each type of air, with respect to the value of the flow rate for standard atmospheric air. The models currently used by the CDT to calculate the density and viscosity of dry and humid air are identified, in addition to proposing new models for the properties of these compositions and for synthetic air. The reference data for the properties of the different air compositions are obtained using the REFPROP software in its version 10.0. As a product of this analysis, it is found that the composition with the maximum property and flow deviation from standard atmospheric air is synthetic air with the maximum nitrogen content, resulting in a 0.6% higher flow value in LFE. Experimental verification of the effect of the composition is also carried out when using dry and synthetic air, in calibrations between laminar pattern and thermal mass meter, as well as reciprocating piston meter. In these calibrations, the En score is used as a statistical criterion to establish if the relative errors of the flow in the different types of air are comparable to each other. | |
dc.description.degreelevel | Pregrado | |
dc.description.degreename | Ingeniero Mecánico | |
dc.format.mimetype | application/pdf | |
dc.identifier.instname | Universidad Industrial de Santander | |
dc.identifier.reponame | Universidad Industrial de Santander | |
dc.identifier.repourl | https://noesis.uis.edu.co | |
dc.identifier.uri | https://noesis.uis.edu.co/handle/20.500.14071/14752 | |
dc.language.iso | spa | |
dc.publisher | Universidad Industrial de Santander | |
dc.publisher.faculty | Facultad de Ingeníerias Fisicomecánicas | |
dc.publisher.program | Ingeniería Mecánica | |
dc.publisher.school | Escuela de Ingeniería Mecánica | |
dc.rights | info:eu-repo/semantics/openAccess | |
dc.rights.accessrights | info:eu-repo/semantics/openAccess | |
dc.rights.coar | http://purl.org/coar/access_right/c_abf2 | |
dc.rights.creativecommons | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
dc.rights.license | Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | |
dc.subject | Humedad | |
dc.subject | Calibration | |
dc.subject | LFE | |
dc.subject | Desecante | |
dc.subject | Flujo de gas | |
dc.subject | Viscosidad | |
dc.subject | Densidad | |
dc.subject | Aire seco | |
dc.subject | Aire sintético | |
dc.subject.armarc | Fluidos | |
dc.subject.keyword | Desiccant | |
dc.subject.keyword | Gas flow | |
dc.subject.keyword | Calibration | |
dc.subject.keyword | Viscosity | |
dc.subject.keyword | Density | |
dc.subject.keyword | Dry air | |
dc.subject.keyword | Synthetic air | |
dc.subject.keyword | Humidity | |
dc.title | Evaluación del Efecto del Uso de Aire Atmosférico, Aire Sintético y la Actualización del Modelado Matemático de las Propiedades del Aire Involucradas Sobre la Calibración de los Principales Medidores de Volumen y Flujo de Gas de la Corporación Centro de Desarrollo Tecnológico del Gas. | |
dc.title.english | Evaluation of the Effect of the Use of Atmospheric Air, Synthetic Air and the Update of the Mathematical Modeling of the Air Properties Involved in the Calibration of the Main Gas Volume and Flow Meters of the Development Center of Gas Technology Corporation. | |
dc.type.coar | http://purl.org/coar/resource_type/c_7a1f | |
dc.type.hasversion | http://purl.org/coar/version/c_b1a7d7d4d402bcce | |
dc.type.local | Tesis/Trabajo de grado - Monografía - Pregrado |
Files
Original bundle
1 - 3 of 3
No Thumbnail Available
- Name:
- Nota de proyecto.pdf
- Size:
- 370.72 KB
- Format:
- Adobe Portable Document Format
No Thumbnail Available
- Name:
- Carta de autorizacion de uso.pdf
- Size:
- 156.65 KB
- Format:
- Adobe Portable Document Format
License bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- license.txt
- Size:
- 2.18 KB
- Format:
- Item-specific license agreed to upon submission
- Description: